- Operating frequency range: from 85 MHz to 4.8 GHz (typical 5.4 GHz)
- Measured parameters: S11, cable loss
- Impedance: 50Ω
- Number of ports: 1
- Dynamic range: not less than 97 dB
- Measurement time at one frequency: 200 µs
- Output power: -30; -10 dB/mW
- Measurement of S11, |S21|, |S12|, S22 (when using 2 reflectometers)
- Connection to devices without measurement cable
- The smallest and lightest among the CABAN series reflectometers
Certificate of Approval of Type of Measuring Instruments by the Russian Federation No. 55466
1-year warrantyNeed a consult?
Our experts will answer any technical questionThe vector reflectometer (vector network analyzer) CABAN R54 is designed to measure the magnitude and phase of the reflection coefficient, the magnitude of the transmission coefficient (when using 2 reflectometers), cable loss, in the operating frequency range from 85 MHz to 4.8 GHz. This vector reflectometer can be connected directly to the device under test without the need for a microwave measurement cable, eliminating measurement errors. As a result, high reliability of operation and calibration stability are ensured.
Application
If you would like to consult with our engineers regarding the application of the CABAN R54 vector reflectometer in your project, please use the "Ask a Question" form.
Software
The CABAN R54 vector reflectometer comes with RVNA software compatible with the Windows operating system. Upon additional request, device software for operation in the Linux OS is provided. The software can be installed on a computer, laptop, or tablet, which in turn is connected to the reflectometer via a USB interface.
Working Frequency Range | from 85 MHz to 4.8 GHz (typical 5.4 GHz) |
Number of Measurement Ports | 1 |
Type of Measurement Port Connectors | N, plug |
Wave Impedance of Measurement Ports, Ohm | 50 |
Measured Parameters | S11, cable loss S11, |S21|, |S12|, S22 (when using 2 devices) |
Output Power, dBm | -30; -10 |
Dynamic Range*, at least, dB | 97 |
Number of Measurement Points per Sweep | from 2 to 100,001 |
Minimum Measurement Time at One Frequency, μs | 200 |
Minimum Frequency Step Size, Hz | 10 |
Measurement Filter Bandwidth (with 1/3 coefficient), Hz | from 10 to 30,000 |
Power Supply | from USB interface |
Operating Ambient Temperature Range, °C | from -10 to +50 |
Dimensions (L×W×H), mm, not exceeding | 120×43×23 |
Weight, kg, not exceeding | 0.25 |
* with an intermediate frequency filter bandwidth of 100 Hz ```
Name | Quantity, pcs |
Vector Reflectometer | 1 |
USB Cable | 1 |
Software | 1 |
User Manual | 1 |
Verification Procedure | 1 |
Form | 1 |
Notes:
1. The specific modification of the vector reflectometer is determined upon ordering.
2. Accessories, including measurement adapters and calibration tools, are supplied upon separate order.
```